- AUW analytical balances, maximum weight range 120 g to 320 g
- AUX analytical balances, maximum weight range 120 g to 320 g
- AUY analytical balances, maximum weight range 120 g to 220 g
AU Series balances inc...
Description
AUW and AUW-D Model Analytical Balances
Model | AUW120D | AUW220D | AUW120 | AUW220 | AUW320 |
---|---|---|---|---|---|
Capacity | 120 g/42 g | 220 g/82 g | 120 g | 220 g | 320 g |
Minimum display | 0.1 mg/0.01 mg | 0.1 mg | |||
Repeatability (standard deviation) |
≤0.1 mg/≤0.02 mg | ≤0.1 mg/≤0.05 mg | ≤0.1 mg | ≤0.15 mg | |
Linearity | ±0.2 mg/±0.03 mg | ±0.2 mg/±0.1 mg | ±0.2 mg | ±0.3 mg | |
Response time | 3 secs/12 secs | 3 secs/15 secs | 3 secs | ||
UniBloc weighing technology | Included | ||||
Built-in calibration | Included | ||||
Built-in clock | Included | ||||
GLP/GMP/ISO calibration report | Included | ||||
Analog bar graphic display | Included | ||||
Specific gravity measurement mode | Included | ||||
Below weight measurement | Included | ||||
Formulation mode | Included | ||||
Internal Timer Output | Included | ||||
Operating temperature range | 5 to 40⁰C | ||||
Sensitivity stability against temperature range1 | ±2 ppm/⁰C (10 to 30⁰C) | ||||
Pan size | Approx. Dia. 80 mm | ||||
Body dimensions W x D x H |
Approx. 220 x 330 x 310 mm | ||||
Weight | Approx. 7.0 kg | ||||
Backlight display | Not Available | Included | |||
Input/output terminal | RS232 I/F |
AUX and AUY Model Analytical Balances
Model | AUX120 | AUX220 | AUX320 | AUY120 | AUY220 |
---|---|---|---|---|---|
Capacity | 120 g | 220 g | 320 g | 120 g | 220 g |
Minimum display | 0.1 mg | ||||
Repeatability (standard deviation) |
≤0.1 mg | ≤0.15 mg | ≤0.1 mg | ||
Linearity | ±0.2 mg | ±0.3 mg | ±0.2 mg | ||
Response time | 3 secs | ||||
UniBloc weighing technology | Included | ||||
Built-in calibration | Included | Not available | |||
Built-in clock | Included | Not available | |||
GLP/GMP/ISO calibration report | Included | Not available | |||
Analog bar graphic display | Included | ||||
Specific gravity measurement mode | Included | ||||
Below weight measurement | Included | ||||
Formulation mode | Included | ||||
Internal Timer Output | Included | Not available | |||
Operating temperature range | 5 to 40⁰C | ||||
Sensitivity stability against temperature range1 | ±2 ppm/⁰C (10 to 30⁰C) | ||||
Pan size | Approx. Dia. 80 mm | ||||
Body dimensions W x D x H |
Approx. 220 x 330 x 310 mm | ||||
Weight | Approx. 7.0 kg | ||||
Backlight display | Not Available | ||||
Input/output terminal | RS232 I/F |